Calibration Laboratory
| Devices |
Range |
Expanded uncertainty (k=2) |
Method |
Multimeters Voltmeters Ampermeters Ohmmeters |
220 mV .. 1100 V DC |
> (0.001 % + 5 µV) |
Calibrator 5720A |
| 2.2 mV .. 1100 V AC (10 Hz .. 1 MHz) |
> (0.008 % + 10 µV) |
| 1 Ω .. 100 MΩ |
> (0.002 % + 50 µΩ) |
| 220 µA .. 2.2 A DC |
> (0.001 % + 15 nA) |
| 220 µA .. 2.2 A AC (10 Hz .. 10 kHz) |
> (0.02 % + 15 nA) |
Calibrators Resitors |
100 mV .. 1000 V DC |
> (0.001 % + 5 µV) |
Multimeter 1281 |
| 100 mV .. 1100 V AC (10 Hz .. 1 MHz) |
> (0.01 % + 10 µV) |
| 10 Ω .. 1 GΩ |
> (0.002 % + 50 µΩ) |
| 100 µA .. 1 A DC |
> (0.001 % + 15 nA) |
| 100 µA .. 1 A AC (10 Hz .. 5 kHz) |
> (0.02 % + 15 nA) |
Power meters Energy meters (active, reactive) |
30 .. 500 V AC |
> 0.01 % |
AP Standards |
| 1 mA .. 250 A AC |
| 0 .. 360 - phase angle |
| 40 .. 75 Hz basic frequency |
| Sub-harmonics and harmonics |
|
|
|
|
| © 2011 Applied Precision Ltd. |